These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of SNX-2112, a selective Hsp90 inhibitor, in plasma samples by high-performance liquid chromatography and its application to pharmacokinetics in rats.
    Author: Zhai QQ, Gong GQ, Luo Y, Wang QD, Xia M, Xing GW, Li YC, Jiang JH, Liu Z, Liu QY, Wang YF.
    Journal: J Pharm Biomed Anal; 2010 Dec 01; 53(4):1048-52. PubMed ID: 20675090.
    Abstract:
    A sensitive and specific reversed-phase high-performance liquid chromatography method with ultraviolet detection has been developed and validated for the identification and quantification of SNX-2112 in rat plasma. Following sample preparation using liquid-liquid extraction, the analytes were separated by the mobile phase acetonitrile-water (40:60, v/v) with an Agilent RP-HPLC column (ZORBAX SB-C18, 5 microm, 4.6 mm x 250 mm) at a flow rate of 1 ml/min, column temperature of 30 degrees C and detection wavelength of 251 nm. The retention time of SNX-2112 was 11.2 min. A good linear relationship was obtained in the concentration range studied (0.07-21 microg/ml, R(2)>0.9982), and the LLOD and LLOQ for SNX-2112 were 0.02 and 0.07 microg/ml, respectively. The mean absolute recovery of SNX-2112 in plasma ranged from 88.58 to 99.61% at the studied concentrations. The intra- and inter-batch relative standard deviations were 1.7-3.5 and 1.9-4.4%, respectively. This method was successfully applied to pharmacokinetic studies in rats after intravenous administration of SNX-2112.
    [Abstract] [Full Text] [Related] [New Search]