These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Noise measurement from magnitude MRI using local estimates of variance and skewness.
    Author: Rajan J, Poot D, Juntu J, Sijbers J.
    Journal: Phys Med Biol; 2010 Aug 21; 55(16):N441-9. PubMed ID: 20679694.
    Abstract:
    In this note, we address the estimation of the noise level in magnitude magnetic resonance (MR) images in the absence of background data. Most of the methods proposed earlier exploit the Rayleigh distributed background region in MR images to estimate the noise level. These methods, however, cannot be used for images where no background information is available. In this note, we propose two different approaches for noise level estimation in the absence of the image background. The first method is based on the local estimation of the noise variance using maximum likelihood estimation and the second method is based on the local estimation of the skewness of the magnitude data distribution. Experimental results on synthetic and real MR image datasets show that the proposed estimators accurately estimate the noise level in a magnitude MR image, even without background data.
    [Abstract] [Full Text] [Related] [New Search]