These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A miniaturized spatial temperature gradient capillary electrophoresis system with radiative heating and automated sample introduction for DNA mutation detection.
    Author: Xu ZR, Li Q, Fan XF, Zhang HD, Fang J.
    Journal: Electrophoresis; 2010 Sep; 31(18):3137-43. PubMed ID: 20734373.
    Abstract:
    A miniaturized spatial temperature gradient CE system with automated sample introduction for DNA mutation detection was established. Continuous electrokinetic sample injection was achieved by combining an automated slotted-vial array sample introduction device to the spatial temperature gradient CE system. The temperature gradient was produced by a radiative heating system with a single graphite block heater, and the stability of the temperature gradient was investigated. The temperature variation of each measure point was 0.12-0.21% RSD (n=7) within 6 h. A 14-cm Teflon AF-coated silica capillary was used both as the separation channel and as the liquid-core waveguide tube of fluorescence signal. Under a temperature gradient from 54.8 to 59.5°C, a low range control mutation standard (209 bp) was separated within 4 min with only 5.6 nL sample consumption. Automated continuous sample introducing and changing were realized with a carryover of 3.3%. Utility of the system was further demonstrated by detecting K-ras gene mutations in paraffin tissue sections from two colorectal cancer patients.
    [Abstract] [Full Text] [Related] [New Search]