These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Automated chemiluminescent dual-analyte immunoassay based on resolved immunosensing channels.
    Author: Yang Z, Zong C, Yan F, Ju H.
    Journal: Talanta; 2010 Sep 15; 82(4):1462-7. PubMed ID: 20801356.
    Abstract:
    A novel system of series-wound immunosensing channels (SWIC) was proposed for automated chemiluminescent (CL) dual-analyte immunoassay by immobilizing respectively different capture antibodies on the inner walls of series-wound glass channels. This system could use a single enzyme as label to perform multiplex immunoassay in one fluid way. Using alpha-fetoprotein (AFP) and carcinoembryonic antigen (CEA) as model analytes, the mixture including AFP, horseradish peroxidase (HRP)-labeled anti-AFP antibody, CEA and HRP-labeled anti-CEA antibody was introduced into the SWIC for carrying out the on-line incubation. Upon injection of CL substrate the CL signals from the two immunosensing channels were conveniently resolved and near-simultaneously collected with the aid of optical shutter. AFP and CEA could be rapidly assayed in the ranges of 1.0-100 and 1.0-80 ng/ml with detection limits of 0.41 and 0.39 ng/ml, respectively. The assay results of clinical serum samples were in an acceptable agreement with the reference values. This designed flow-through immunosensing system based on SWIC provided an automated, reusable, simple, sensitive and low-cost approach for multianalyte immunoassay.
    [Abstract] [Full Text] [Related] [New Search]