These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Feasibility of FAIR imaging for evaluating tumor perfusion. Author: Cho JH, Cho G, Song Y, Lee C, Park BW, Lee CK, Kim N, Park SB, Kang JS, Kang MR, Kim HM, Kim YR, Cho KS, Kim JK. Journal: J Magn Reson Imaging; 2010 Sep; 32(3):738-44. PubMed ID: 20815076. Abstract: PURPOSE: To evaluate the feasibility of flow-sensitive alternating inversion recovery (FAIR) for measuring blood flow in tumor models. MATERIALS AND METHODS: In eight mice tumor models, FAIR and dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) was performed. The reliability for measuring blood flow on FAIR was evaluated using the coefficient of variation of blood flow on psoas muscle. Three regions of interest (ROIs) were drawn in the peripheral, intermediate, and central portions within each tumor. The location of ROI was the same on FAIR and DCE-MR images. The correlation between the blood flow on FAIR and perfusion-related parameters on DCE-MRI was evaluated using the Pearson correlation coefficient. RESULTS: The coefficient of variation for measuring blood flow was 9.8%. Blood flow on FAIR showed a strong correlation with Kep (r = 0.77), percent relative enhancement (r = 0.73), and percent enhancement ratio (r = 0.81). The mean values of blood flow (mL/100 g/min) (358 vs. 207), Kep (sec(-) (1)) (7.46 vs. 1.31), percent relative enhancement (179% vs. 134%), and percent enhancement ratio (42% vs. 26%) were greater in the peripheral portion than in the central portion (P < 0.01). CONCLUSION: As blood flow measurement on FAIR is reliable and closely related with that on DCE-MR, FAIR is feasible for measuring tumor blood flow.[Abstract] [Full Text] [Related] [New Search]