These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Long-term clinical and hemodynamic performance of the Hancock II versus the Perimount aortic bioprostheses.
    Author: Chan V, Kulik A, Tran A, Hendry P, Masters R, Mesana TG, Ruel M.
    Journal: Circulation; 2010 Sep 14; 122(11 Suppl):S10-6. PubMed ID: 20837899.
    Abstract:
    BACKGROUND: The Medtronic Hancock II and the Carpentier-Edwards Perimount are among the world's most commonly used aortic bioprostheses. However, a direct comparison of their clinical performance is lacking. To minimize biases inherent to between-center comparisons, we examined these prostheses within a large, contemporary, single-center cohort. METHODS AND RESULTS: Between 1990 and 2007, 1659 patients (mean age, 73.1±9.3 years) underwent aortic valve replacement with either the Hancock II (N=1021) or the Perimount (N=638). Patients were prospectively followed-up with serial clinic visits and echocardiograms for up to 16 years (mean, 5.0±3.3 years). There was no significant difference in aortic root size preoperatively (P=0.7). Aortic root enlargement was more commonly performed with the Perimount (P<0.001), and the manufacturer valve size of the implanted prosthesis was larger with the Hancock II (P<0.001). Postoperatively, peak and mean transprosthesis gradients were higher for the Hancock II (32.7±0.7 and 16.0±0.3 mm Hg, respectively) than for the Perimount (24.9±0.7 and 13.4±0.4 mm Hg, respectively; P<0.001). However, no difference in left ventricular mass regression was observed at late follow-up (P=0.9). Unadjusted 10-year survival was 59.4%±2.4% for the Hancock II and 70.2%±3.8% for the Perimount (P=0.07). Multivariable predictors of survival did not include prosthesis type (P=0.2). CONCLUSIONS: For the same manufacturer valve size, the Perimount is larger, which may warrant enlarging the aortic root more often, and it is associated with better hemodynamics than the Hancock II. These differences do not impact survival or left ventricular mass regression, and the long-term clinical performances of the Hancock II and Perimount bioprostheses are equivalent.
    [Abstract] [Full Text] [Related] [New Search]