Title: The ENEA-IRP thoron calibration facility. Author: Calamosca M, Penzo S. Journal: Radiat Prot Dosimetry; 2010 Oct; 141(4):468-72. PubMed ID: 20846973. Abstract: To check the consistency of a (220)Rn measurement, performed by passive alpha track detector (ATD), the use of an accurate (220)Rn exposure calibration facility is mandatory. The ENEA Radon Service developed a new CR-39 ATD-Tn, coupled to the radon ATD-Rn and, to assess its sensitivity, had to design a small calibration facility, which has been recently modified to improve the spatial homogeneity exposure conditions inside the chamber, to get a continuous monitoring of the (220)Rn concentration and to reduce radon contamination. A better knowledge of the circuit response allowed selecting the best-operating conditions and how to regulate the thoron concentration. The active thoron monitor has been changed to serve as a continuous sampling and measuring device rather than a grab one; particular care has been devoted to assess the (216)Po losses on the device's inner surfaces and to evaluate the chamber transit time correction factor.[Abstract] [Full Text] [Related] [New Search]