  • Title: Prototype fiber-optic-based pressure probe with built-in temperature compensation with signal recovery by coherence reading.
    Author: Rao YJ, Jackson DA.
    Journal: Appl Opt; 1993 Dec 01; 32(34):7110-3. PubMed ID: 20856576.
    Abstract:
    A prototype multimode fiber-based Fabry-Perot interferometric pressure probe utilizing a corrugated diaphragm with built-in temperature compensation, with signal recovery by coherence reading, has been constructed and demonstrated. A separate fiber-optic-based temperature sensor was incorporated into the pressure sensor to permit the pressure measurement to be corrected for the temperature dependence of the pressure probe. A measurement range to resolution of 3.6 × 10(4) - 1 and an overall measurement accuracy of ±0.15% have been achieved. This system represents a practical approach for industrial use.