These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Prototype fiber-optic-based pressure probe with built-in temperature compensation with signal recovery by coherence reading. Author: Rao YJ, Jackson DA. Journal: Appl Opt; 1993 Dec 01; 32(34):7110-3. PubMed ID: 20856576. Abstract: A prototype multimode fiber-based Fabry-Perot interferometric pressure probe utilizing a corrugated diaphragm with built-in temperature compensation, with signal recovery by coherence reading, has been constructed and demonstrated. A separate fiber-optic-based temperature sensor was incorporated into the pressure sensor to permit the pressure measurement to be corrected for the temperature dependence of the pressure probe. A measurement range to resolution of 3.6 × 10(4) - 1 and an overall measurement accuracy of ±0.15% have been achieved. This system represents a practical approach for industrial use.[Abstract] [Full Text] [Related] [New Search]