These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Study on several systems used for chemiluminescent determination of horseradish peroxidase].
    Author: Yang X, Wu L, Chen D, Zen Y.
    Journal: Hua Xi Yi Ke Da Xue Xue Bao; 1990 Sep; 21(3):293-7. PubMed ID: 2093069.
    Abstract:
    Four systems were studied used for chemiluminescent determination of HRP and the determination conditions optimized. In luminol-H2O2 (NaOH) system, HRP was determined conveniently in the range of 0.05-4 pmol, with a correlation coefficient of 0.998 for the linear log-log equation and a limit of detection of 16 fmol HRP. In pyrogallol-H2O2 (pH 6.5) system, 0.1-100 pmol of HRP were determined, based on three log-log regression equations with correlation coefficients of 0.997, 0.994 and 0.999, respectively, and limits of detection of 3-17 fmol HRP. Luminol-H2O2-para iodophenol system was used to determine HRP in the range to 10-50 fmol, based on the chemiluminescence intensity at 1 min and the correlation coefficient of the log-log equation being 0.996 with a limit of detection of 10 fmol HRP. When p-hydroxybiphenyl was used to take the place of p-iodophenol and the chemiluminescence intensity at 1 min or 3 min measured, 6-40 fmol HRP were determined with a correlation coefficient of 0.994 or 0.995 for the log-log equation and a limit of detection of 6.5 or 3.3 fmol HRP.
    [Abstract] [Full Text] [Related] [New Search]