  • Title: Carbon nanotube composite coated platinum electrode for detection of Ga(III).
    Author: Abbaspour A, Khoshfetrat SM, Sharghi H, Khalifeh R.
    Journal: J Hazard Mater; 2011 Jan 15; 185(1):101-6. PubMed ID: 20980100.
    Abstract:
    This study demonstrates the application of composite multi-walled carbon nanotube (MWCNT) polyvinylchloride (MWNT-PVC) based on 7-(2-hydroxy-5-methoxybenzyl)-5,6,7,8,9,10-hexahydro-2H benzo [b][1,4,7,10,13] dioxa triaza cyclopentadecine-3,11(4H,12H)-dione ionophore for gallium sensor. The sensor shows a good Nernstian slope of 19.68 ± 0.40 mV/decade in a wide linear range concentration of 7.9 × 10(-7) to 3.2 × 10(-2)M of Ga(NO(3))(3). The detection limit of this electrode is 5.2 × 10(-7)M of Ga(NO(3))(3). This proposed sensor is applicable in a pH range of 2.7-5.0. It has a short response time of about 10s and has a good selectivity over nineteen various metal ions. The practical analytical utility of this electrode is demonstrated by measurement of Ga(III) in river water.