Title: Evaluation of force requirements during use of recycled plastic syringes. Author: Mwaniki DL. Journal: Trop Geogr Med; 1990 Oct; 42(4):380-4. PubMed ID: 2100082. Abstract: Forces required to withdraw the piston in recycled syringes were investigated. New disposable syringes were used as a reference. The force required in withdrawal of the piston for recycled syringes was between 3 and 8 times that required for the reference syringes. Lubrication reduced the force for the recycled syringes substantially. A modification of syringe recycling protocol is suggested.[Abstract] [Full Text] [Related] [New Search]