  • Title: [The use of a hybrid scanner as a human radiation counter].
    Author: Luchkov AB, Trushin VI.
    Journal: Med Radiol (Mosk); 1990 Jun; 35(6):27-9. PubMed ID: 2113154.
    Abstract:
    The most promising devices of those for radionuclide diagnosis that can be used for control of nuclides incorporated in the human body during a radiation incident are hybrid scanners. The authors made the calibration of devices of this type to register a mixture of 137Cs and 134Cs. The minimum detected activity was 40 nCi. A methodological error of measurements was +/- 3%. A mathematical criterion for optimization of the geometry of measurements was drawn for devices, including the methodological and statistical components of a total error. This criterion was used for finding optimum sizes of a slot collimator of the device.