These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [The use of a hybrid scanner as a human radiation counter]. Author: Luchkov AB, Trushin VI. Journal: Med Radiol (Mosk); 1990 Jun; 35(6):27-9. PubMed ID: 2113154. Abstract: The most promising devices of those for radionuclide diagnosis that can be used for control of nuclides incorporated in the human body during a radiation incident are hybrid scanners. The authors made the calibration of devices of this type to register a mixture of 137Cs and 134Cs. The minimum detected activity was 40 nCi. A methodological error of measurements was +/- 3%. A mathematical criterion for optimization of the geometry of measurements was drawn for devices, including the methodological and statistical components of a total error. This criterion was used for finding optimum sizes of a slot collimator of the device.[Abstract] [Full Text] [Related] [New Search]