These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A new method of analysis of peroxydisulfate using ion chromatography and its application to the simultaneous determination of peroxydisulfate and other common inorganic ions in a peroxydisulfate matrix. Author: Khan NE, Adewuyi YG. Journal: J Chromatogr A; 2011 Jan 21; 1218(3):392-7. PubMed ID: 21185030. Abstract: A new method for the determination of peroxydisulfate using ion chromatography has been developed. Elution of peroxydisulfate was effected by isocratic elution using 200 mM NaOH at 40°C. A modification of the method using gradient elution was able to simultaneously determine other common inorganic ions (nitrate, nitrite, sulfate and chloride) down to significantly low concentrations in a peroxydisulfate matrix. The relative standard deviations (RSD) were in the range of 0.5-5%, for peak areas and <0.2% for peak retention times. The recoveries were between 95% and 120% for a concentration range of about 0.5-42 ppm. The limit of detection for peroxydisulfate ion was 0.2 ppm and for the other ions were ≤2×10(-2) ppm. The calibration curves were linear with slope and intercepts close to 1 and 0, respectively.[Abstract] [Full Text] [Related] [New Search]