These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Dendrimer modified graphite sensors for detection of anticancer drug Daunorubicin by voltammetry and electrochemical impedance spectroscopy.
    Author: Erdem A, Karadeniz H, Caliskan A.
    Journal: Analyst; 2011 Mar 07; 136(5):1041-5. PubMed ID: 21203608.
    Abstract:
    The development of amino-terminated G4 PAMAM dendrimer (PDR) modified disposable electrodes were developed as the first time in our study by using the dendrimer modified disposable graphite (PDR-PGE) and multiwalled carbon nanotube based screen-printed graphite (PDR-MWCNT-SPE) electrodes. Firstly, the microscopic characterization of bare PGEs and PDR modified PGEs was performed. These sensors were then applied for electrochemical monitoring of an anticancer drug, Daunorubicin (DNR). The enhanced oxidation signal of DNR was measured at +0.50 V by using differential pulse voltammetry (DPV) in combination with the PDR-PGEs. The detection limit, estimated from S/N = 3, corresponds accordingly to 317 nM and 128 nM for DNR respectively at the PGE and PDR-PGE. The voltammetric results were consistent with electrochemical impedance spectroscopy (EIS) that was used to characterize the successful modification of PDR onto the surface of PGE and MWCNT-SPE.
    [Abstract] [Full Text] [Related] [New Search]