These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Multi-laboratory evaluation of a scrub typhus diagnostic kit.
    Author: Kelly DJ, Wong PW, Gan E, Chye CT, Cowan D, Lewis GE.
    Journal: Am J Trop Med Hyg; 1990 Sep; 43(3):301-7. PubMed ID: 2121057.
    Abstract:
    Scrub typhus is a major cause of febrile illness throughout the Asia-Pacific region. It is commonly undiagnosed, partly because of the lack of a simple, reliable diagnostic test which can be used in clinical laboratories. The indirect immunoperoxidase technique, configured into a test kit, was provided to technicians who were trained in its use. They used the kit during a 2 year field trial in their respective clinical hospital laboratories throughout Malaysia. In an evaluation using 1,722 consecutive sera tested in those laboratories, the kit was found to have a median sensitivity for IgG detection of 0.85 (range 0.33-0.95), a median specificity of 0.94 (range 0.88-1.00), reproducibility of 0.86, and efficiency of 0.92 when compared to the reference laboratory. In a proficiency survey in which 10 laboratories received 3 coded test samples, all but 2 laboratories had results within 1 dilution of the reference laboratory in quantitating specific IgG, whereas 7 laboratories were within 1 dilution in quantitating IgM. The shelf life of the kit was at least 1 year at 4 degrees C.
    [Abstract] [Full Text] [Related] [New Search]