These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Midterm results of 506 solid trispiked reflection cementless acetabular components for primary total hip arthroplasty.
    Author: Corten K, McCalden RW, Teo Y, Charron KD, MacDonald SJ, Bourne RB.
    Journal: J Arthroplasty; 2011 Dec; 26(8):1350-6. PubMed ID: 21266305.
    Abstract:
    The purpose of this prospective study was to evaluate the outcomes and survivorship of a trispiked, sintered bead-coated titanium shell with a polished inner surface, no screw holes, and an improved locking mechanism. A total of 506 total hip arthroplasties with a minimum 5-year follow-up were available for review at a mean of 7 years (range, 5-11 years). Three sockets (0.6%) were revised for reasons other than aseptic loosening, and 14 (2.8%) polyethylene liners were exchanged. There was no difference in revision rate between non-cross-linked and highly cross-linked liners (P = .4). There were no cases of radiographic loosening. Retroacetabular osteolysis was identified in 2%. The overall 5-year and 10-year Kaplan-Meier survivorship was 97.5% and 97.4%, respectively, whereas survivorship of the shell was 99.8%.
    [Abstract] [Full Text] [Related] [New Search]