These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Suggested baseline ranges for performance testing of the Fujifilm FCR Capsula XL computed radiography system.
    Author: Thiele DL, Schick DK, Barr S.
    Journal: Australas Phys Eng Sci Med; 2010 Dec; 33(4):335-40. PubMed ID: 21286892.
    Abstract:
    Due to the difficulty in obtaining expected baseline values from the supplier for the performance testing of the Fujifilm FCR Capsula XL computed radiography system, performance test results from the acceptance testing of 42 Capsula systems were reviewed. Guidance ranges are provided for nine system performance tests which can be applied to future testing of calibrated systems.
    [Abstract] [Full Text] [Related] [New Search]