Search MEDLINE/PubMed
Title: Suggested baseline ranges for performance testing of the Fujifilm FCR Capsula XL computed radiography system. Author: Thiele DL, Schick DK, Barr S. Journal: Australas Phys Eng Sci Med; 2010 Dec; 33(4):335-40. PubMed ID: 21286892. Abstract: Due to the difficulty in obtaining expected baseline values from the supplier for the performance testing of the Fujifilm FCR Capsula XL computed radiography system, performance test results from the acceptance testing of 42 Capsula systems were reviewed. Guidance ranges are provided for nine system performance tests which can be applied to future testing of calibrated systems.[Abstract] [Full Text] [Related] [New Search]