These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A new solid phase micro extraction for simultaneous head space extraction of ultra traces of polar and non-polar compounds. Author: Alizadeh R, Najafi NM, Kharrazi S. Journal: Anal Chim Acta; 2011 Mar 09; 689(1):117-21. PubMed ID: 21338766. Abstract: The results of the innovative study on a new stationary phase with high efficiency based on ZnO nano and micro rod coating on fused silica are reported in this paper. ZnO nanorods with a diameter in the range of 70-300 nm and the length of about 500 nm, have been grown on fused silica fibers using a hydrothermal process. The extraction properties of the fiber were investigated using headspace solid-phase microextraction (HS-SPME) mode coupled with gas chromatography-mass spectrometry detection (GC-MS) for 1,4-dichloro-nitrobenzene, biphenyl and acenaphthene. The calibration curves were linear up to 10(2)-10(7) ng L(-1) (R(2)>0.995) with detection limits of 10(-3) ng L(-1) for biphenyl and acenaphthene and 10 ng L(-1) for 1,4-dichloro-nitrobenzene. The RSD for single fiber and fiber-to-fiber were less than 7.0 and 11.5%, respectively. The high stability of the ZnO coating is proved at relatively high temperatures (up to 300°C) with a high extraction capacity and long lifespan (more than 100 times). Promising recoveries (91-102%) were obtained in environmental water samples analysis by applying the proposed technique.[Abstract] [Full Text] [Related] [New Search]