Title: Evaluation of the wear performance of a polycarbonate-urethane acetabular component in a hip joint simulator and comparison with UHMWPE and cross-linked UHMWPE. Author: St John K, Gupta M. Journal: J Biomater Appl; 2012 Jul; 27(1):55-65. PubMed ID: 21343216. Abstract: Acetabular hip joint components manufactured from gamma-sterilized ultra high molecular weight polyethylene (UHMWPE), gamma cross-linked UHMWPE, or polycarbonate-urethane (PCU) polymers were evaluated in a hip joint simulator, using cobalt alloy femoral components, for at least 5 million cycles. The volume of material losses due to wear was calculated for each type of sample, based upon mass loss measurements, every 500,000 cycles. The loss of material for the conventional UHMWPE was much higher than for the cross-linked UHMWPE, showing about a 70% reduction in wear due to cross-linking. The material loss for the PCU samples appears to have been at least 24% lower than for the cross-linked UHMWPE. Based upon these results, the PCU material seems to have potential for use as an alternative bearing material to UHMWPE for total hip replacement surgeries.[Abstract] [Full Text] [Related] [New Search]