These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Time-resolved contrast-enhanced magnetic resonance angiography (CEMRA) of the left atrium-pulmonary veins complex with half dose of intravenous gadolinium-based contrast agent. Technical feasibility and comparison with a conventional CEMRA, full contrast dose protocol.
    Author: Faggioni L, Zampa V, Ortori S, Picano E, De Lucia R, Soldati E, Bongiorni MG, Neri E, Bartolozzi C.
    Journal: Eur J Radiol; 2012 Feb; 81(2):250-6. PubMed ID: 21353421.
    Abstract:
    PURPOSE: To evaluate feasibility and image quality of time-resolved contrast-enhanced magnetic resonance angiography (CEMRA) of the left atrium-pulmonary veins (LA-PV) complex with half dose of intravenous gadolinium-based contrast agent (GBCA) in patients candidate to percutaneous radiofrequency ablation of atrial fibrillation. METHODS AND MATERIALS: Fifty-seven patients underwent CEMRA of the LA-PV complex on a 1.5T MRI scanner. On 24/57 patients, a conventional fast-spoiled gradient-echo (FSPGR) CEMRA acquisition was run using 0.2 mL/kg of 0.5M GBCA at 2 mL/s flow rate (protocol A), while in 33/57 patients a time-resolved multiphase CEMRA sequence (Time-Resolved Imaging of Contrast KineticS, TRICKS) was performed after intravenous injection of 0.1 mL/kg of the same GBCA at 3 mL/s flow rate (protocol B). Contrast enhancement was measured in the LA (LAe) and in the PA (PAe), and the LAe/PAe ratio was calculated. Diagnostic quality of Maximum Intensity Projection (MIP), Volume Rendering (VR), and Virtual Endoscopy (VE) reconstructions was also assessed visually using a semiquantitative score. RESULTS: LAe was comparable with both protocols, while PAe was lower with protocol B than with protocol A (p = 0.0217). Moreover, the LAe/PAe ratio was significantly higher with protocol B than with protocol A (p = 0.0044). Finally, image quality of MIP, VR, and VE reconstructions was significantly better with protocol B than with protocol A (p = 0.0005, p = 0.0001, and p = 0.005, respectively). CONCLUSIONS: CEMRA of the LA-PV complex is feasible with TRICKS and half-dose GBCA and yields better separation between the LA-PV complex and the PA, as well as better image quality of MIP, VR, and VE reconstructions than a conventional FSPGR sequence performed with full GBCA dose.
    [Abstract] [Full Text] [Related] [New Search]