These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Miniature endoscopic optical coherence tomography probe employing a two-axis microelectromechanical scanning mirror with through-silicon vias. Author: Liu L, Wu L, Sun J, Lin E, Xie H. Journal: J Biomed Opt; 2011 Feb; 16(2):026006. PubMed ID: 21361690. Abstract: We present the design and experimental results of a new MEMS-based endoscopic optical coherence tomography (OCT) probe. The uniqueness of this miniature OCT imaging probe is a two-axis MEMS mirror with through-silicon vias (TSVs) for interconnecting. The TSV interconnection enables ultracompact probe design, successfully reducing the probe size to only 2.6 mm in diameter. The MEMS mirror is actuated by an electrothermal actuator that is capable of scanning ± 16° at only 3.6 V DC. Two-dimensional and three-dimensional OCT images of microspheres embedded in PDMS and acute rat brain tissue have been obtained with this miniature probe in a time-domain OCT system.[Abstract] [Full Text] [Related] [New Search]