  • Title: Evaluation of expanded polytetrafluoroethylene-covered stents for the treatment of venous outflow stenosis in hemodialysis access grafts.
    Author: Chan MG, Miller FJ, Valji K, Kuo MD.
    Journal: J Vasc Interv Radiol; 2011 May; 22(5):647-53. PubMed ID: 21371910.
    Abstract:
    PURPOSE: To investigate the use of an expanded polytetrafluoroethylene-coated nitinol stent-graft in patients with failing hemodialysis access circuits and compare grafts with and without a heparin-coated surface. MATERIALS AND METHODS: A single-institution retrospective review of 16 patients (average age, 55 years; seven men) treated for hemodialysis access circuit stenoses was performed. Patients had a prosthetic arteriovenous (AV) graft (n = 10) or autogenous AV fistula (n = 6), all with clinically relevant access stenoses that were refractory to angioplasty alone; 10 patients also had access circuit thrombosis at the time of treatment. Stents were placed in the basilic (n = 4), brachial (n = 7), cephalic (n = 4), and femoral (n = 1) veins. Study endpoints included primary treatment area patency, primary circuit patency, primary assisted circuit patency, and secondary circuit patency, followed to a maximum of 6 months. RESULTS: Primary treatment area patency rates at 30 days and 180 days were 75% and 31%, respectively. Secondary patency rates at the same time points were 88% and 68%, respectively. Mean primary treatment area patency was 93 days. Thrombosis was a major cause of failure, occurring in seven of 11 cases of primary treatment area patency loss. Subgroup analysis demonstrated that the heparin-coated configuration (n = 7) showed a 6-month primary patency rate of 57%, compared with 11% for the stent-grafts without heparin bonding (n = 9; P = .06). CONCLUSIONS: Overall, the majority of deployed stent-grafts lost primary patency within 6 months, with thrombosis a major cause of failure. However, the addition of a heparin-coated surface appeared to improve results in these otherwise treatment-refractory access circuits.
    [Abstract] [Full Text] [Related] [New Search]