These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of expanded polytetrafluoroethylene-covered stents for the treatment of venous outflow stenosis in hemodialysis access grafts. Author: Chan MG, Miller FJ, Valji K, Kuo MD. Journal: J Vasc Interv Radiol; 2011 May; 22(5):647-53. PubMed ID: 21371910. Abstract: PURPOSE: To investigate the use of an expanded polytetrafluoroethylene-coated nitinol stent-graft in patients with failing hemodialysis access circuits and compare grafts with and without a heparin-coated surface. MATERIALS AND METHODS: A single-institution retrospective review of 16 patients (average age, 55 years; seven men) treated for hemodialysis access circuit stenoses was performed. Patients had a prosthetic arteriovenous (AV) graft (n = 10) or autogenous AV fistula (n = 6), all with clinically relevant access stenoses that were refractory to angioplasty alone; 10 patients also had access circuit thrombosis at the time of treatment. Stents were placed in the basilic (n = 4), brachial (n = 7), cephalic (n = 4), and femoral (n = 1) veins. Study endpoints included primary treatment area patency, primary circuit patency, primary assisted circuit patency, and secondary circuit patency, followed to a maximum of 6 months. RESULTS: Primary treatment area patency rates at 30 days and 180 days were 75% and 31%, respectively. Secondary patency rates at the same time points were 88% and 68%, respectively. Mean primary treatment area patency was 93 days. Thrombosis was a major cause of failure, occurring in seven of 11 cases of primary treatment area patency loss. Subgroup analysis demonstrated that the heparin-coated configuration (n = 7) showed a 6-month primary patency rate of 57%, compared with 11% for the stent-grafts without heparin bonding (n = 9; P = .06). CONCLUSIONS: Overall, the majority of deployed stent-grafts lost primary patency within 6 months, with thrombosis a major cause of failure. However, the addition of a heparin-coated surface appeared to improve results in these otherwise treatment-refractory access circuits.[Abstract] [Full Text] [Related] [New Search]