  • Title: Highly sensitive fluorescence detection system for microfluidic lab-on-a-chip.
    Author: Ryu G, Huang J, Hofmann O, Walshe CA, Sze JY, McClean GD, Mosley A, Rattle SJ, deMello JC, deMello AJ, Bradley DD.
    Journal: Lab Chip; 2011 May 07; 11(9):1664-70. PubMed ID: 21431240.
    Abstract:
    We demonstrate a compact, low cost and practical fluorescence detection system for lab-on-a-chip applications. The system comprises a commercially available InGaN light emitting diode (501 nm) as light source, an organic or silicon photodiode detector, absorptive dye coated colour filters and linear and reflective polarisers. An injection moulded polystyrene microfluidic chip is used as the platform for fluorescence immunoassays for cardiac markers myoglobin and CK-MB. The optical limit of detection (LOD) is measured using a TransFluoSphere® suspension at 5.6 × 10(4) beads µl(-1) which can be equated to ∼3 nM fluorescein equivalent concentration. The LOD for the human plasma immunoassays is measured as 1.5 ng ml(-1) for both myoglobin and CK-MB.