These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The usefulness of an earphone-type infrared tympanic thermometer during cardiac surgery with cardiopulmonary bypass: clinical report.
    Author: Masamune T, Yamauchi M, Wada K, Iwashita H, Okuyama K, Ino H, Yamakage M, Ishiyama T, Matsukawa T.
    Journal: J Anesth; 2011 Aug; 25(4):576-9. PubMed ID: 21519927.
    Abstract:
    We evaluated the usefulness of a novel earphone-type infrared tympanic thermometer (IRT) during cardiac surgery with cardiopulmonary bypass. Tympanic membrane temperature (T(Tym)) was monitored using the IRT inserted into the right ear canal of 12 adult patients (ASA III) who had been scheduled for elective cardiac surgery with cardiopulmonary bypass under general anesthesia. Rectum (T(Rec)) and nasopharyngeal temperatures (T(Naso)) were also monitored, and all temperatures were recorded at 5-min intervals during cardiopulmonary bypass. Operating room temperature was kept at 20°-27°C; a conductive warming/cooling system was used to control the patient's body temperature. Of 265 measurements obtained, body temperature range was 31.6°-37.6°C. No complications were related to site of insertion of the monitoring probe. Significant correlations were seen between T(Tym) and T(Naso) (r = 0.971, P < 0.001), and T(Tym) and T(Rec) (r = 0.759, P < 0.001). A Bland-Altman plot showed that average temperature of T (Tym) was 0.06°C above T(Naso) (±0.66°C, 2 SD) and 0.12°C below T(Rec) (±1.78°C, 2 SD). We conclude that an earphone-type IRT is noninvasive and hygienic and could continuously evaluate selective cerebral temperature during cardiopulmonary bypass in adults.
    [Abstract] [Full Text] [Related] [New Search]