These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A 2:1 AV rhythm: an adverse effect of a long AV delay during DDI pacing and its prevention by the ventricular intrinsic preference algorithm in DDD mode. Author: Minamiguchi H, Oginosawa Y, Kohno R, Tamura M, Takeuchi M, Otsuji Y, Abe H. Journal: Pacing Clin Electrophysiol; 2012 Jul; 35(7):e189-92. PubMed ID: 21671954. Abstract: A 91-year-old woman received a dual-chamber pacemaker for sick sinus syndrome and intermittently abnormal atrioventricular (AV) conduction. The pacemaker was set in DDI mode with a 350-ms AV delay to preserve intrinsic ventricular activity. She complained of palpitation during AV sequential pacing. The electrocardiogram showed a 2:1 AV rhythm from 1:1 ventriculoatrial (VA) conduction during ventricular pacing in DDI mode with a long AV interval. After reprogramming of the pacemaker in DDD mode with a 250-ms AV interval and additional 100-ms prolongation of the AV interval by the ventricular intrinsic preference function, VA conduction disappeared and the patient's symptom were alleviated without increasing unnecessary right ventricular pacing.[Abstract] [Full Text] [Related] [New Search]