Search MEDLINE/PubMed
Title: The impact of extra-column band broadening on the chromatographic efficiency of 5 cm long narrow-bore very efficient columns. Author: Fekete S, Fekete J. Journal: J Chromatogr A; 2011 Aug 05; 1218(31):5286-91. PubMed ID: 21726868. Abstract: Small columns packed with core-shell and sub-2 μm totally porous particles and monolith columns are very popular to conduct fast and efficient chromatographic separations. In order to carry out fast separations, short (2-5 cm) and narrow-bore (2-2.1 mm) columns are used to decrease the analyte retention volume. Beside the column efficiency, another significant issue is the extra-column band-spreading. The extra-column dispersion of a given LC system can dramatically decrease the performance of a small very efficient column. The aim of this study was to compare the extra-column peak variance contribution of several commercially available LC systems. The efficiency loss of three different type 5 cm long narrow bore, very efficient columns (monolith, sub-2 μm fully porous and sub-2 μm core-shell packing) as a function of extra-column peak variance, and as a function of flow rate and also kinetic plots (analysis time versus apparent column efficiency) are presented.[Abstract] [Full Text] [Related] [New Search]