These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development of a method for the determination of naphthalene and phenanthrene in workplace air using diffusive sampling and thermal desorption GC-MS analysis.
    Author: Lindahl R, Claesson AS, Khan MA, Levin JO.
    Journal: Ann Occup Hyg; 2011 Jul; 55(6):681-7. PubMed ID: 21742628.
    Abstract:
    Diffusive sampling methods have been validated for the determination of naphthalene and phenanthrene in workplace air. The diffusive sampler tested was the Perkin Elmer ATD tube, and the analysis was performed with thermal desorption, gas chromatography, and mass spectrometric detection. The sampling methods were validated in controlled test atmospheres, mainly according to the protocol proposed in the European standard EN 838. For the determination of naphthalene, the diffusive sampling rate was 0.41 ml min(-1) with a coefficient of variation (CV) of 19%. The mean sampling rate for phenanthrene was 0.49 ml min(-1) with a CV of 21%. Field tests confirmed the naphthalene results but could not be used to confirm the phenanthrene results. The method is not recommended for phenanthrene sampling unless the method has been tested in the specific environment and the results confirm the laboratory tests.