These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Quantifying the mechanical properties of human skin to optimise future microneedle device design. Author: Groves RB, Coulman SA, Birchall JC, Evans SL. Journal: Comput Methods Biomech Biomed Engin; 2012; 15(1):73-82. PubMed ID: 21749225. Abstract: Microneedle devices are a promising minimally invasive means of delivering drugs/vaccines across or into the skin. However, there is currently a diversity of microneedle designs and application methods that have, primarily, been intuitively developed by the research community. To enable the rational design of optimised microneedle devices, a greater understanding of human skin biomechanics under small deformations is required. This study aims to develop a representative stratified model of human skin, informed by in vivo data. A multilayer finite element model incorporating the epidermis, dermis and hypodermis was established. This was correlated with a series of in-vivo indentation measurements, and the Ogden material coefficients were optimised using a material parameter extraction algorithm. The finite element simulation was subsequently used to model microneedle application to human skin before penetration and was validated by comparing these predictions with the in-vivo measurements. Our model has provided an excellent tool to predict micron-scale human skin deformation in vivo and is currently being used to inform optimised microneedle designs.[Abstract] [Full Text] [Related] [New Search]