These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Low latency breathing frequency detection and monitoring on a personal computer.
    Author: Leskovsek M, Ahlin D, Cancer R, Hosta M, Enova D, Pusenjak N, Bunc M.
    Journal: J Med Eng Technol; 2011; 35(6-7):322-9. PubMed ID: 21767133.
    Abstract:
    We demonstrate a low latency respiratory/breathing frequency detection system that is fast (<5 ms), easy to operate, requires no batteries or external power supply and operates fully via computer-standard USB connection. Exercises in controlling ones breathing frequency, usually referred to as paced-breathing exercises, have shown positive effects in treating pulmonary diseases, cardiovascular diseases and stress/anxiety-related disorders. We developed a breathing frequency detection system which uses two pairs of microphones to detect exhalation activity, eliminate noise from the environment and stream the recording data via USB connection to a personal computer. It showed 97.1% reliability (10 subjects) when monitoring breathing activity in non-guided free breathing and 100% reliability (10 subjects) when monitoring breathing activity during interactive paced-breathing exercises. We also evaluated the breathing frequency detection systems noise elimination functionality which showed a reduction of 84.2 dB for stationary (white noise) and a reduction of 79.3 dB for non-stationary (hands clapping) noise.
    [Abstract] [Full Text] [Related] [New Search]