Search MEDLINE/PubMed


  • Title: Quantitation of human chorionic gonadotropin with the Abbott IMx immunochemistry analyzer.
    Author: van Oers RJ, Leerkes B, Bertschi I.
    Journal: J Clin Chem Clin Biochem; 1990 Jan; 28(1):59-63. PubMed ID: 2179460.
    Abstract:
    The analytical performance of the Abbott IMx automated system was studied and initial results from the human chorionic gonadotropin assay are presented: Within run and between day precision was good, the majority of all CVs lying between 3 and 5%; Patient results showed excellent comparability with our routine method; No significant specimen-related carry-over was detected; The automated IMx dilution protocol was found to be superior to the manual procedure; The usefulness of the instrument in routine use and as a "Stat" analyzer was very good.