These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Preferred performance of the high-performance membrane in the case of online hemodiafiltration.
    Author: Kawanishi H.
    Journal: Contrib Nephrol; 2011; 173():36-43. PubMed ID: 21865774.
    Abstract:
    High-performance membrane (HPM) dialyzers have become commonly used in online hemodiafiltration (ol-HDF). HPM dialyzers with a markedly improved function of removing high-molecular solutes have appeared, but excessive albumin loss often occurs using these. Higher albumin loss was observed when these protein-permeable dialyzers were used for post-dilution ol-HDF. When albumin loss is regarded as the criterion, dialyzers with a 0.5 g or lower albumin loss in the hemodialysis mode should be used for post-dilution HDF, whereas those with a 2 g or greater albumin loss should be selected for pre-dilution HDF. The performance of recent HPM dialyzers has been markedly developed, and equivalent removal efficiency can now be achieved with non-ol-HDF. To demonstrate the efficacy of ol-HDF, it is necessary to compare hemodialysis and ol-HDF using a type V dialyzer (β(2)-microglobulin clearance ≥70 ml/min) classified in Japan, and re-investigate the significance of removing high-molecular solutes larger than β(2)-microglobulin and protein-bound solutes.
    [Abstract] [Full Text] [Related] [New Search]