These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Measurements of relative depth doses and Cerenkov light using a scintillating fiber-optic dosimeter with Co-60 radiotherapy source. Author: Jang KW, Yoo WJ, Moon J, Han KT, Park JY, Lee B. Journal: Appl Radiat Isot; 2012 Jan; 70(1):274-7. PubMed ID: 21889353. Abstract: In this study, we fabricated a scintillating fiber-optic dosimeter, which consists of an organic scintillator and a plastic optical fiber, for radiotherapy dosimetry. To select an adequate kind and length of scintillator for γ-rays generated from a Co-60 source, scintillating light from various kinds and lengths of organic scintillators is measured. Using a scintillating fiber-optic dosimeter, the γ-rays generated from a Co-60 therapy unit are measured and relative doses are obtained according to the field size of the γ-ray beam and the depth in a water phantom. Also, Cerenkov light generated by the interactions of primary or secondary electrons and the plastic optical fiber is measured with different field sizes and depths of a water phantom using a background optical fiber.[Abstract] [Full Text] [Related] [New Search]