These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Novel conductometric biosensor based on three-enzyme system for selective determination of heavy metal ions.
    Author: Soldatkin OO, Kucherenko IS, Pyeshkova VM, Kukla AL, Jaffrezic-Renault N, El'skaya AV, Dzyadevych SV, Soldatkin AP.
    Journal: Bioelectrochemistry; 2012 Feb; 83():25-30. PubMed ID: 21903487.
    Abstract:
    A differential pair of planar thin-film interdigitated electrodes, deposited on a ceramic pad, was used as a conductometric transducer. The three-enzyme system (invertase, mutarotase, glucose oxidase), immobilized on the transducer surface, was used as a bioselective element. The ratio between enzymes in the membrane was found experimentally considering the highest biosensor sensitivity to substrate (sucrose) and heavy metal ions. Optimal concentration of sucrose for inhibitory analysis was 1.25 mM and incubation time in the investigated solution amounted to 10-20 min. The developed biosensor demonstrated the best sensitivity toward ions Hg(2+) and Ag(+). A principal possibility of the biosensor reactivation either by EDTA solution after inhibition with silver ions or by cysteine solution after inhibition with mercury ions was shown.
    [Abstract] [Full Text] [Related] [New Search]