  • Title: Autoradiography using storage phosphor technology.
    Author: Johnston RF, Pickett SC, Barker DL.
    Journal: Electrophoresis; 1990 May; 11(5):355-60. PubMed ID: 2194789.
    Abstract:
    We describe the application of photostimulable storage phosphor imaging plates to autoradiography of samples labeled with 32P, 14C and 35S. Imaging plates can detect a 32P-labeled sample at an exposure level of 1 disintegration/mm2, and 14C and 35S at 25 disintegrations/mm2. Compared to X-ray film, imaging plates offer increased sensitivity of 15- to 250-fold for 32P and 20- to 100-fold for 14C and 35S. Resolution is in the 0.3 mm range and is therefore suitable for any gel or blot application.