These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Lepirudin dose-dependently increases thrombelastography parameters at therapeutic plasma concentrations as measured with ROTEM® - a pilot study. Author: Feuring M, Wehling M, Schultz A. Journal: Int J Clin Pharmacol Ther; 2011 Oct; 49(10):626-8. PubMed ID: 21961488. Abstract: BACKGROUND: The aim of this in-vitro pilot study was to assess the usefulness of the thrombelastograph ROTEM® for determining the anticoagulant activity of lepirudin. METHODS: The ROTEM® parameters, clotting-time, clot formation time and maximum clot firmness were measured in the presence of increasing concentrations of lepirudin (10-4 μg/ml - 10 μg/ml). Citrated blood was obtained from 16 healthy male subjects. RESULTS: Clotting-time increased from 79.1 ± 53.4 s at baseline to 194.1 ± 151.9 s at a drug concentration of 1μg/ml as measured with EXTEM (p < 0.0001). Borderline significance was found for the difference between maximum clot firmness at baseline (60.2 ± 4.3 mm) and after drug application (55.5 ± 6.5 mm). CONCLUSIONS: This pilot investigation shows that the ROTEM® device may be suitable for monitoring lepirudin at low concentrations but the results should be confirmed in a larger study and the ROTEM® device validated against standard methods.[Abstract] [Full Text] [Related] [New Search]