These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Implementation of the elemental library stratified sampling technique on the GUI-based Monte Carlo library least squares (MCLLS) approach for EDXRF analysis.
    Author: Li F, Gardner RP.
    Journal: Appl Radiat Isot; 2012 Jul; 70(7):1243-9. PubMed ID: 22001412.
    Abstract:
    The Monte Carlo-based Energy Dispersive X-ray Fluorescence (MCXRF) elemental analyzer approach can be used to accurately calculate elemental weight fractions in rock formations or metal alloys, by least-squares regression of the measured X-ray spectrum with computer-generated elemental library spectra. The analog sampling scheme in a Monte Carlo simulation cannot generate satisfactory results for trace elements due to their low sampling probability. A stratified sampling variance-reduction technique is proposed and implemented to enable each element in the measured sample to have equal opportunity of being sampled with adjusted weights. Essentially, all computed elemental libraries have good statistical precision and improve the accuracy of the elemental analysis results.
    [Abstract] [Full Text] [Related] [New Search]