Title: Design of a manual device to measure ankle joint stiffness and range of motion. Author: Kobayashi T, Leung AK, Hutchins SW. Journal: Prosthet Orthot Int; 2011 Dec; 35(4):478-81. PubMed ID: 22013156. Abstract: BACKGROUND AND AIM: Ankle joint stiffness and its range of motion (ROM) are commonly assessed to determine the appropriate mechanical characteristics required in an effective ankle-foot orthosis (AFO) prescription. The aim of this technical note is to present the design of a manual device that enables their convenient measurement in the clinical setting and to demonstrate its reliability. TECHNIQUE: The manual device was designed with a torquemeter, a potentiometer, a steering wheel, a rotary plate, and a foot plate. The measurement of resistive torque at 0° (neutral), 5° (dorsiflexion) and 10° (dorsiflexion) ankle angular positions demonstrated the high reliability of the device with Intraclass Correlation Coefficient (ICC) (1,k) values over 0.97. DISCUSSION: Quantitative measurement of ankle joint stiffness and ROM by this manual device would provide objective information that could potentially assist AFO prescriptions. A future study should investigate how to incorporate the measurement obtained from the device into the prescription of an AFO.[Abstract] [Full Text] [Related] [New Search]