These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Ppb-level detection of nitric oxide using an external cavity quantum cascade laser based QEPAS sensor. Author: Dong L, Spagnolo V, Lewicki R, Tittel FK. Journal: Opt Express; 2011 Nov 21; 19(24):24037-45. PubMed ID: 22109428. Abstract: Geometrical parameters of micro-resonator for a quartz enhanced photoacoustic spectroscopy sensor are optimized to perform sensitive and background-free spectroscopic measurements using mid-IR quantum cascade laser (QCL) excitation sources. Such an optimized configuration is applied to nitric oxide (NO) detection at 1900.08 cm(-1) (5.26 µm) with a widely tunable, mode-hop-free external cavity QCL. For a selected NO absorption line that is free from H(2)O and CO(2) interference, a NO detection sensitivity of 4.9 parts per billion by volume is achieved with a 1-s averaging time and 66 mW optical excitation power. This NO detection limit is determined at an optimal gas pressure of 210 Torr and 2.5% of water vapor concentration. Water is added to the analyzed mixture in order to improve the NO vibrational-translational relaxation process.[Abstract] [Full Text] [Related] [New Search]