These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Paclitaxel coating of the luminal surface of hemodialysis grafts with effective suppression of neointimal hyperplasia. Author: Baek I, Bai CZ, Hwang J, Nam HY, Park JS, Kim DJ. Journal: J Vasc Surg; 2012 Mar; 55(3):806-814.e1. PubMed ID: 22226184. Abstract: OBJECTIVES: Paclitaxel coating of hemodialysis grafts is effective in suppressing neointimal hyperplasia in the graft and vascular anastomosis sites. However, paclitaxel can have unwanted effects on the surrounding tissues. To reduce such problems, we developed a method to coat the drug only on the luminal surface of the graft, with little loading on the outer surface. METHODS: A peristaltic pump and a double-solvent (water and acetone) system were used to achieve an inner coating of paclitaxel. At the ratio of 90% acetone, paclitaxel was homogeneously coated only on the luminal surface of the graft without changing the physical properties. To determine its effect, grafts were implanted between the common carotid artery and the external jugular vein in pigs using uncoated control grafts (n = 6) and low-dose (n = 6, 0.22 μg/mm(2)) and high-dose (n = 6, 0.69 μg/mm(2)) paclitaxel inner-coated grafts. Cross-sections of graft-venous anastomoses were analyzed histomorphometrically 6 weeks after placement to measure the patency rate, percentage of luminal stenosis, and neointimal area. RESULTS: No signs of infection or bacterial contamination were observed in the paclitaxel inner-coated groups. Only one of the six control grafts was patent, but all of the paclitaxel-coated grafts were patent, with little neointima. The mean ± standard error values of percentage luminal stenosis were 75.7% ± 12.7% (control), 17.5% ± 3.1% (low dose), and 19.7% ± 3.0% (high dose). The values for the neointimal area (in mm(2)) were 8.77 ± 1.66 (control), 3.53 ± 0.73 (lose dose), and 4.24 ± 0.99 (high dose). Compared with the control group, paclitaxel inner-coated vascular grafts significantly suppressed neointimal hyperplasia (low dose, P = .001; high dose, P = .002). Myofibroblast proliferation and migration into the graft interstices confirmed the firm attachment of the implanted graft to the surrounding tissue. CONCLUSIONS: Paclitaxel coating on the inner luminal surface of vascular grafts was effective in suppressing neointimal hyperplasia, with little inhibition of myofibroblast infiltration within the graft wall.[Abstract] [Full Text] [Related] [New Search]