  • Title: Rapid, sensitive, and multiplexed on-chip optical sensors for micro-gas chromatography.
    Author: Reddy K, Guo Y, Liu J, Lee W, Oo MK, Fan X.
    Journal: Lab Chip; 2012 Mar 07; 12(5):901-5. PubMed ID: 22245960.
    Abstract:
    We developed and characterized a rapid, sensitive and integrated optical vapor sensor array for micro-gas chromatography (μGC) applications. The sensor is based on the Fabry-Pérot (FP) interferometer formed by a micrometre-thin vapor-sensitive polymer layer coated on a silicon wafer. The thickness and the refractive index of the polymer vary in response to the vapor analyte, resulting in a change in the reflected intensity of the laser impinged on the sensor. In our study, four different polymers were coated on four wells pre-etched on a silicon wafer to form a spatially separated sensor array. A CMOS imager was employed to simultaneously monitor the polymers' response, thus enabling multiplexed detection of a vapor analyte passing through the GC column. A sub-second detection time was demonstrated. In addition, a sub-picogram detection limit was achieved, representing orders of magnitude improvement over the on-chip vapor sensors previously reported.
    [Abstract] [Full Text] [Related] [New Search]