These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Electromagnetic analysis of an RF rectangular resonant cavity applicator for hyperthermic treatment using whole-body voxel human model of Japanese adult male. Author: Tange Y, Yabumoto K, Kanai Y. Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2011; 2011():337-40. PubMed ID: 22254317. Abstract: The numerical whole-body voxel human model (numerical model) developed by National Institute of Information and Communications Technology (NICT) was assumed and hyperthemic treatment using radio-frequency wave was investigated. We assumed 51 different human tissues and organs with 2-mm spatial resolution in the numerical model, inserted it into the resonant cavity applicator, and Maxwell's equations were solved by FDTD method with variable mesh. We obtained the realistic energy patterns for a deep-seated tumor as compared to those obtained in our previous studies.[Abstract] [Full Text] [Related] [New Search]