These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of the analytical performances of the novel Beckman Coulter AU5800. Author: Lippi G, Dipalo M, Musa R, Avanzini P, Ferrarini C, Pattini A, Aloe R. Journal: Clin Biochem; 2012 Apr; 45(6):502-4. PubMed ID: 22296728. Abstract: OBJECTIVES: To assess the analytical performance of Beckman Coulter AU5800. DESIGN AND METHODS: Imprecision, linearity and comparison studies were performed on 12 routine clinical chemistry tests. RESULTS: Imprecision was comprised between 0.4 and 3.0%. The linearity was excellent, with regression coefficients always >1.000. The correlation with analogous tests on Unicell DxC ranged from 0.894 and 1.000. The productivity was 4884 tests/h. CONCLUSIONS: We found optimal balance between performance and productivity for the AU5800.[Abstract] [Full Text] [Related] [New Search]