These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Calibration of stack monitors for measurement of noble gases in nuclear facilities.
    Author: Kovar P, Dryak P, Suran J, Gudelis A.
    Journal: Appl Radiat Isot; 2012 Sep; 70(9):2127-9. PubMed ID: 22424839.
    Abstract:
    In nuclear facilities stack monitors are used for the measurement of the volumetric activity of noble gases. Spectrometric measurement is needed because the content of stack effluents is always a mixture of radionuclides. In some nuclear power plants new types of monitors were installed based on HPGe detectors. For efficiency calibration a standard with the radionuclide Xe-127 was developed and calibration curve constructed in the energy range 81 keV-1293 keV. Experiental efficiencies were checked using an MC model.
    [Abstract] [Full Text] [Related] [New Search]