These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Scatter fractions in AMBER imaging. Author: Chotas HG, Floyd CE, Dobbins JT, Lo JY, Ravin CE. Journal: Radiology; 1990 Dec; 177(3):879-80. PubMed ID: 2244003. Abstract: Images of two phantoms were obtained with use of an advanced multiple-beam equalization radiography system, and scatter fractions were estimated with use of a photostimulable phosphor imaging system. Scatter fractions in the equalized images were lower in the mediastinum-equivalent areas and higher in the lung-equivalent areas, relative to images that were conventionally acquired with use of an antiscatter grid. The differences are attributed to a reduction in incident exposure in the lungs and the presence of cross-scatter between lung and mediastinal regions.[Abstract] [Full Text] [Related] [New Search]