These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Different manufacturers of syringes: a new source of variability in blood gas, acid-base balance and related laboratory test?
    Author: Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth G, Guidi GC.
    Journal: Clin Biochem; 2012 Jun; 45(9):683-7. PubMed ID: 22440459.
    Abstract:
    OBJECTIVE: Evaluate whether four different brands/types of heparin coated syringes can represent a source of variability in blood gas analysis (BGA). DESIGN AND METHODS: Blood was collected from one hundred volunteers into different syringes: Syringe I (lithium heparin and calcium balanced); Syringe II: in-house prepared (sodium heparin); Syringe III: (spray-dried calcium-balanced lithium heparin); Syringe IV (lyophilized electrolyte-balanced lithium heparin). RESULTS: Significant differences were as follows: a) Syringe I vs II: pO(2), sO(2), pCO(2)(t), cHCO(3)(-), ctCO(2), base excess (BE), total hemoglobin (tHb), sodium (Na(+)), potassium (K(+)), calcium (Ca(2+)), glucose (Glu), lactate (Lac), O(2) Hb and p 50; b) Syringe I vs III: pH, pO(2), cHCO(3)(-), ctCO(2), BE, Na(+), Glu, Lac and p 50; c) Syringe I vs IV: pH, pO(2), sO(2), pCO(2)(t), BE, Na(+), K(+), Ca(2+), Glu, Lac and O(2) Hb; d) Syringe II vs III: pH, pO(2), sO(2), pCO(2)(t), cHCO(3)(-), ctCO(2), ctO(2), tHb, Na(+), K(+), Ca(2+), Lac and p 50; e) Syringe II vs IV: pH, pO(2), sO(2), pCO(2)(t), cHCO(3)(-), ctCO(2), BE, tHb, Na(+), K(+), Ca(2+), Lac, O(2) Hb and p 50; f) Syringe III vs IV: pH, pO(2), sO(2), cHCO(3)(-), ctCO(2), ctO(2), BE, Na(+), K(+), Ca(2+), O(2) Hb and p 50. CONCLUSION: The different manufacturers of syringes can represent new source of variability on BGA.
    [Abstract] [Full Text] [Related] [New Search]