These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Near-patient potassium and sodium measurements: evaluation of ion-selective electrodes (Ionometer EF2) in a dialysis department.
    Author: Madsen HH, Antonsen S, Nielsen HK.
    Journal: Blood Purif; 1990; 8(3):171-6. PubMed ID: 2244994.
    Abstract:
    For potassium and sodium monitoring, an Ionometer EF2 was installed in our dialysis unit. The apparatus measured linearly, and between-series and within-series variations were comparable to flame photometry. No difference in performance was disclosed when operated by one or several persons. The 95% limits of agreement between Ionometer EF2 and flame photometry were for potassium -0.29 and 0.43 mmol/l and for sodium -5.0 and 2.89 mmol/l, respectively. Contrary to sodium, the accidental difference for potassium was greater than expected from analytical variation.
    [Abstract] [Full Text] [Related] [New Search]