These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Quality control of blood glucose self-monitoring using the Glucosignal test strip and Haemo-Glucotest 20-800].
    Author: Schimke E, Ratzmann KP.
    Journal: Z Gesamte Inn Med; 1990 Aug 01; 45(15):444-6. PubMed ID: 2247992.
    Abstract:
    In 157 capillary blood samples of type I diabetics parallel determinations of the blood sugar concentration were performed by means of the visual evaluation of the test strips Glucosignal and Haemo-Glucotest 20-800 in comparison to an enzymatic reference method by trained special personnel. On the basis of the correlation coefficients the two test strips showed a good accordance between visual evaluation and the reference method (r = 0.98 and 0.97, respectively). In comparison to the reference method in Glucosignal there is a tendency to underestimation (-0.67 mmol/l), on the other hand in the Haemo-Glucotest 20-800 the blood glucose concentrations are on an average +0.87 mmol/l above the reference value. In visual valuation of the blood glucose concentrations by means of Haemo-Glucotest 20-800 with increasing blood glucose concentration an increasing inexactness is to be stated, whereas Glucosignal is characterized by more favourable parameters of quality.
    [Abstract] [Full Text] [Related] [New Search]