Title: A piezoelectric immunosensor for chloramphenicol detection in food. Author: Karaseva NA, Ermolaeva TN. Journal: Talanta; 2012 May 15; 93():44-8. PubMed ID: 22483874. Abstract: A piezoelectric immunosensor on the basis of electrogenerated polymer for detecting trace quantities of chloramphenicol, a drug which is used in veterinary for treatment and prevention of infectious diseases, was developed. The detection was conducted in a competitive format with the use of hapten-protein conjugate and monoclonal antibodies. The limit of detection is 0.2 ng ml(-1), the calibration curve is linear in the range of concentrations 0.5-100.0 ng ml(-1). Chloramphenicol was detected in food (meat, milk, egg, honey).[Abstract] [Full Text] [Related] [New Search]