These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Early recognition of an entrapped pulmonary artery catheter by blood leaking into the syringe and thermistor connector during cardiac surgery.
    Author: Chan WH, Hsu CH, Lu CC, Tsai CS, Yang HY, Liaw WJ, Lin TC.
    Journal: Acta Anaesthesiol Taiwan; 2012 Mar; 50(1):38-40. PubMed ID: 22500913.
    Abstract:
    An 84-year-old male was scheduled for coronary artery bypass graft surgery under general anesthesia. During cardiopulmonary bypass, the leakage of blood into the syringe being used for balloon inflation and the thermistor connector of the pulmonary artery catheter (PAC) was detected. Resistance was encountered when trying to withdraw the PAC. A surgical suture of the right atrium cannulation was stitched to PAC and was immediately released. Early detection of surgical damage to PAC and recognition of the entrapped PAC by gently withdrawing it avoided possible life-threatening complications, including pulmonary air embolism, and the inevitable of resternotomy. Transesophageal echocardiography, chest radiography, and fluoroscopy can help confirm any postoperative surgical damage following closure of the sternum or while in the intensive care unit.
    [Abstract] [Full Text] [Related] [New Search]