These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Enhancing pedicle screw fixation in the aging spine with a novel bioactive bone cement: an in vitro biomechanical study. Author: Zhu Q, Kingwell S, Li Z, Pan H, Lu WW, Oxland TR. Journal: Spine (Phila Pa 1976); 2012 Aug 01; 37(17):E1030-7. PubMed ID: 22531472. Abstract: STUDY DESIGN: A paired biomechanical study of pedicle screws augmented with bone cement in a human cadaveric and osteoporotic lumbar spine model. OBJECTIVES.: To evaluate immediate strength and stiffness of pedicle screw fixation augmented with a novel bioactive bone cement in an osteoporotic spine model and compare it with polymethylmethacrylate (PMMA) cement. SUMMARY OF BACKGROUND DATA: A novel bioactive bone cement, containing nanoscale particles of strontium and hydroxyapatite (Sr-HA), can promote new bone formation and osteointegration and provides a promising reinforcement to the osteoporotic spine. Its immediate mechanical performance in augmenting pedicle screw fixation has not been evaluated. METHODS: Two pedicle screws augmented with Sr-HA and PMMA cement were applied to each of 10 isolated cadaveric L3 vertebrae. Each screw was subjected to a toggling test and screw kinematics were calculated. The pedicle screw was subjected to a pullout test until failure. Finally, the screw coverage with cement was measured on computed tomographic images. RESULTS: Screw translations in the toggling test were consistently larger in the Sr-HA group than in the PMMA group (1.4 ± 1.2 mm vs. 1.0 ± 1.1 mm at 1000 cycles). The rotation center was located closer to the screw tip in the Sr-HA group (19% of screw length) than in the PMMA group (37%). The only kinematic difference between Sr-HA and PMMA cements was the screw rotation at 1000 cycles (1.5° ± 0.9° vs. 1.3° ± 0.6°; P = 0.0026). All motion parameters increased significantly with more loading cycles. The pullout force was higher in the PMMA group than the Sr-HA group (1.40 ± 0.63 kN vs. 0.93 ± 0.70 kN), and this difference was marginally significant (P = 0.051). Sr-HA cement covered more of the screw length than PMMA cement (79 ± 19% vs. 43 ± 19%) (P = 0.036). CONCLUSION: This paired-design study identified some subtle but mostly nonsignificant differences in immediate biomechanical fixation of pedicle screws augmented with the Sr-HA cement compared with the PMMA cement.[Abstract] [Full Text] [Related] [New Search]